Two adjustable 5-mm diameter f/2 quartz collimating lenses
Cell compartment that accepts 10-cm cuvettes
Built-in filter slot
Manual light block for dark measurements
Internal channels designed to accept a constant-temperature water source (for heating and cooling of the base and cuvette, by means of convection)
SMA 905 terminations for coupling to fiber optics.
Specifications
Path length
10 cm
Collimating lenses (UV-VIS-NIR)
Quartz lenses optimized from 200 nm-2 ¥ìm
5 mm diameter
f/2
Collimating lens termination
SMA 905
Sample compartment dimensions
2.0" x 1.5" (LW)
Filter slot
Accepts filters up to ¨ù" (6 mm) in thickness
Base material
Aluminum
Base length
5.5"
Connections in base
Water inlet/outlet
Water input fittings
1/8" NPT pipe thread
Recommended optical fibers*
200 ¥ìm illumination fiber; 50 ¥ìm read fiber
Z dimension
15 mm
* There is no single combination of optical fibers that satisfies the requirements of every application. As a rule, however, it is best to use a large-diameter (>50 ¥ìm) illumination fiber to get maximum light throughput and a small-diameter (<50 ¥ìm) read fiber to achieve the best optical resolution